April 30th, 2020 - The Performance of CeBr3 Detectors

Berkeley Nucleonics offers a wide range of custom and standard scintillators including Sodium Iodide, Cesium Iodide, and Cerium Bromide.

High-resolution scintillation detectors are now offering improved performance with lower intrinsic background. Standard cerium bromide (CeBr3) detectors are now readily available from BNC (up to 3x3 inch encapsulated sizes) with greatly improved sensitivity. This paper discusses the merits of CeBr3 from the standpoint of energy resolution and reduction of intrinsic background from actinium-227 (227Ac) contamination.

Detector Resolution

CeBr3 detectors have several advantages over lanthanum bromide (LaBr3) detectors even though LaBr3 is widely accepted as the highest resolution scintillation detector. LaBr3 boasts 3% resolution at 662 keV and CeBr3 has a resolution of about 4% at 662 keV. This allows both detectors to achieve distinct separation of the 609 keV peak of bismuth-214 (214Bi) from the 662 keV peak of cesium-137 (137Cs) (which cannot be accomplished with sodium iodide (NaI(Tl)) detectors). Generally, the ability to resolve spectra at high energies can be accomplished quite well with NaI(Tl). Resolving lower energy spectra of uranium (U) and plutonium (Pu) is important, and it is in this lower energy region that the resolution of CeBr3 outperforms that of LaBr3 [1]. This is noted when observing the plutonium-239 (239Pu) spectrum as shown in Fig. 1. The primary 239Pu peaks (332, 375 and 414 keV) stand out clearly with CeBr3. Also, the Xrays at ~105 keV and the 201 keV gamma line are more pronounced with CeBr3. The low energy lines of weapons-grade uranium are also better resolved with CeBr3. At even lower energies the resolution of LaBr3 continues to fall off dramatically such that the resolution of the 60 and 26 keV energy line of americium-241 (241Am) is worse than NaI(Tl).

Intrinsic Background

The typical background for CeBr3 detectors is primarily from 227Ac (the parent nucleus is uranium-235 (235U)). This contribution to background from the 227Ac contamination is approximately 0.02 counts/sec/cm3 [2]. The concentration of 227Ac atoms per cerium (Ce) atoms is of the order of 4x10-16. Even though this is a small amount of contamination, it does represent a significant contribution to the background, especially in the 1.2 to 2.2 MeV region of the energy spectrum as seen in Fig. 2. Ac, Ce, and lanthanum (La) are all chemically homologous elements and therefore, they are extremely difficult to separate one from another. Through the selection of material being screened before processing, the 227Ac background can now be essentially eliminated (about 0.001 counts/sec/cm3 [2]). Also seen in Fig. 2 is the much higher background in the LaBr3 detectors. In addition to the 227Ac there is significant contamination from lanthanum-138 (138La) which is responsible for a very high background near 32 keV and continuing up to the 1461 keV region. Another region of background in LaBr3 begins at about 1520 keV through 2800 keV. This is from natural uranium (primarily the 214Bi progeny) giving CeBr3 a sensitivity greater than LaBr3 by an order of magnitude in this region of the spectrum. The reason this is
so important is that detection of weapons-grade Pu and weapons-grade U relies on the detection of the 2615 keV energy line to verify its presence. Depending on the activity and the distance to the source, a longer acquisition may be necessary for verification. Therefore, it is not only important to have the highest sensitivity at this energy (CeBr3 improves it by a factor of 10) but with LaBr3, high background in this region could mislead first-line responders with a false positive for weapons-grade material.

References

[1] Comparison of LaBr3: Ce and NaI(Tl) Scintillators for Radio-Isotope Identification Devises, B. D. Milbrath, et. al., Pacific Northwest National Laboratory (PNNL)

[2] P. Schotanus, et al., Nuclear Instruments and Methods in Physics Research A 729 (2013) 596-204