Model DB-2 | NIM Random Pulse Generator

You are here:

Model DB-2 | NIM Random Pulse Generator

The Model DB-2 is a pulse generator which accurately simulates the random and pile-up characteristics of pulses from a radiation detector. It provides pulses which are mono-energetic over a broad range of average count rates. Under high count rate conditions, the pulses will pile up as shown in the diagram below. This characteristics is useful in determining pile-up or count-rate effects and in measuring the resolution of high count rate spectroscopy systems. Besides random pulse, the Model DB-2 also provides mono energetic repetitive pulses. In this mode, the DB-2 is an excellent general purpose pulser. When the EXT REF input is used with an external ramp, the DB-2 provides sliding pulses to quickly check system or component linearity.

Overview

Features
  • Random Tail pulses for MCA, PHA Pile-up and Dead-time Testing
  • Continuously Adjustable Rates from 10 Hz to 1 MHz

Specifications

Rep Rate (Count Rate)

10 Hz to 1 MHz [ Continuously Adjustable]

Mode

Random or Repetitive

Random Distribution

Poisson for intervals greater than 1.4 µs

Pulse Shape

Tail Pulse

Rise Time of Outputs (10-90%)

0.1-20 µs ( 8 steps )

Decay Time Constant (100-37%)

5-1000 µs ( 8 steps )
Rise and Decay time independent of each other for Decay Time/ Rise Time > 10.

Pulse Amplitude (STEP) Characterization

Amplitude Shift with Count Rate: < 0.05% from 10 Hz to 100 kHz
Jitter (resolution): 0.01% RMS
Temperature Coefficient: 0.02%/˚C

Frequency Jitter (Repetitive Mode)

< 0.1%

External Trigger Input

Requires 1 V Positive Pulse
1 kΩ input impedance

Trigger Output

3 V 100 ns Pulse Width
20 ns rise time
50 Ω Output impedance

Output Amplitude Ranges

Repetitive: 0 to 10 V 
Random: 0 to 1 V 
Adjustable by 10-turn Potentiometer from zero to maximum.

Normalize

40% to 100% of Amplitude via 10-Turn Potentiometer

Pulse Output Polarity

Positive or Negative

Attenuation

4 Step Attenuators [x2, x5, x10, and x10] combination to make maximum of x1000

External Reference Input

0 to 10 VDC; 10kΩ Input impedance

Power Requirements

+ 24 V at 65 mA – 24 V at 65 mA +12V at 140 mA -12 V at 40 mA *See BNC Model AP-3 for portable applications

Dimension

Double-width NIM module; 2.70″ wide by 8.70″ high in accordance with TID-20893 (Rev. 3)

Weight

3.5 lbs. net; 7 lbs. shipping

Shipping Dimensions

18x12x9″

Shipping Weight

10 lbs