The Model 090 Optical Module is used with the BNC Model 6040 Optical Pulse Generator Mainframe, much like Model 065 or Model 085. The primary difference with this module is that the spectral peak is 650nm.

Overview

Features
  • Trigger Internal 0.01 Hz to 100 MHz; external 0 to 100 MHz
  • Pulse Width 3 ns to 640 s
  • Impulse 400 ps typical (fixed)
  • Delay 0 ns to 640 s (plus fixed delay)
Applications
  • Photo Multiplier
  • Photo Detector
  • Fiber Optics Testing

Specifications

LIGHT OUT
Spectral Peak 904nm +/- 15nm (nominal)
Spectral Width less than 5nm (nominal)
Power Levels 0 to 2 mW baseline, 100 uW to 2 mW peak (Custom up to 10mW); independently adjustable. Impulse output fixed at 100 uW baseline, 1.5 mW peak (typical).
Resolution 10 uW, 100 uW to 2 mW pulses; 0 to 2 mW in External Modulation and CW.
Transition Times (20% to 80%) 500 ps rise time, 1.5 ns fall time (pulses with zero baseline) 1 ns rise time, 1 ns fall time (pulses with non zero baseline).
Dynamic Range 13 dB, pulses with zero baseline; 20 dB, pulses with nonzero baselineExternal Modulation and CW.
Accuracy (Nonzero Baseline) Absolute: +/- 1 dB, 200 uW to 2 mW. Relative: +/- 0.5 dB, 200 uW to 2 mW pulses;+ 30 uW, 200 uW to 2 mW for External Modulation and CW.
Temperature Coefficient 0.05 dB/°C.
Fiber 6/125 Single Mode Fiber standard (Contact Factory for custom); 0.21 NA.
Connector ST Standared (Contact Factory for custom)
TIMING
Trigger Internal 0.01 Hz to 100 MHz; external 0 to 100 MHz.
Pulse Width 3 ns to 640 s.
Impulse 400 ps typical (fixed).
Delay 0 ns to 640 s (plus fixed delay).
External Drive 0 to 200 MHz, pulses with zero baseline; 0 to 300 MHz, pulses with nonzerobaseline; 300 mV min. amplitude; 50 ohm input.
External Modulation 100 Hz to 700 MHz (-3 dB); 100mV rms/mW sensitivity (typical); 50 ohminput.
Module Delay 10 ns typical (fixed) mainframe's Pulse Out to Light Out. Jitter 100ps rms.
STORE

A ROM determines the instrument's allowable operating conditions andunits of display applicable for the installed module.
A nonvolatile memory (RAM) in each module stores up to ten complete settings of operating conditions (trigger level, slope, etc.) and output pulse parameters

RECALL

The RAM-stored settings can be manually or bus transferred to the mainframeto produce the pre determined outputs.

WEIGHT

2 Ibs net (0.9 kg); 5 Ibs shipping (2.3 kg). (When installed in a Model 6040 Mainframe, est. 24 lbs. shipping)

AMBIENT

Same as Model 6040 mainframe.